31 Prima 451 – User Manual
User Maintenance
6.2 A200SP
Absorber
6.2.1
Removing
the
Absorber
A
ss
e
m
b
l
y
WARNING
Remove the machine from the MRI facility before
commencing any maintenance or repair procedure.
After service or repair work, check that all tools, test/
monitoring equipment, and unused / replaced components
are removed from the machine before it is returned to the
MRI facility.
C
A
U
TIO
N
The absorber assembly weighs approximately 15 kg (when
empty). R
emoval and refitting must only be carried out by
qualified service personnel. When the
absorber is lifted or
carried by hand, always support the weight of the unit under
the base.
Do not lift the absorber by gripping any of the components
attached to the manifold block (1).
C
A
U
TIO
N
The canister (2) may contain condensate. Before removing
the absorber from the pole-mount assembly, refer to the
A200SP user manual for instructions on draining the canister
and disposing of absorbent.
1. Set the brakes on the anaesthetic machine castors.
2. Refer to the A200SP user manual and disconnect all
cable connectors and hoses.
3. Slacken the knob (3), and lift the absorber assembly
from the pole-mount.
2
1
3