Prima 451 – User Manual 23
Installation and Pre-use Checks
5.3
V
a
po
r
i
ze
r
s
Penlon Sigma Delta MRI vaporizers can be supplied as part of
the Prima 451 MRI system (see Appendix 2).
5.3.1 Pre-use Checks -
V
a
po
r
i
ze
r
s
On ALL vaporizers, before use:
1. Check all joints for gas tightness.
2. Check vaporizer agent level.
CAUTION
See section 3.7.4 for additional information for refilling
a key-fill vaporizer in the MR facility.
3. Check for correct agent delivery concentrations - use an
agent analyser.
Vaporizer leak check
NOTE
The machine low pressure test (section 5.2.4) incorporates a
vaporizer leak test.
Always
follow
the
procedures
and
c
h
e
ck
li
s
t
given in
the
instruction
manual
supplied
w
i
t
h
the
vaporizer,
particularly
when
filling
t
h
e
vaporizer
with
anaesthetic
a
g
e
n
t
.
5.3.2 General
I
n
f
o
r
m
a
t
i
on
W
A
RN
I
N
G
Vaporizers
must always be mounted,
n
e
v
e
r
used
f
r
ee-
s
t
a
nd
i
ng
.
Free
standing vaporizers
may be a
cc
i
d
e
n
t
a
ll
y
tipped
resulting
in
excessive
and
uncalibrated volumes
of
anaesthetic
drug
entering
t
h
e
breathing
sys
t
e
m
.
Do not
install
or
connect
any
vaporizer
of a
n
y
description
between the CGO and
t
h
e
breathing
system, unless it is
specifically
designed
for such use.
(This allows
t
h
e oxygen flush flow to pass
through
t
h
e
vaporizer,
and severe
overdosage
may
r
e
s
u
l
t
)
.
5.3.3 Selectatec
Mounting
S
ys
t
e
m
To install the vaporizer:
a) Carefully offer the vaporizer up to the manifold.
b) Check that the gas connection ports on the vaporizer are
aligned with the valves on the manifold.
c) Carefully lower the vaporizer onto the manifold and lock
the vaporizer into position by clockwise rotation of the
locking lever (1) through 90
o
.
NOTE
Do not use excessive force to lock the vaporizer onto the
manifold. Damage to the locking fastener will result.
C
A
U
TIO
N
To prevent damage to the locking shaft (2), ensure that the
gas connection ports are aligned with the valves on the
manifold, and are correctly engaged, before tightening the
locking lever.
5.3.4 Selectatec C
o
m
p
a
t
i
b
l
e MRI
Vaporizers
with
I
n
t
e
r
l
o
ck
W
A
RN
I
N
G
Only MRI
vaporizers
with the
S
e
l
e
c
t
a
t
e
c
compatible
interlock function
will
interlock
i
f
installed
on a two
s
t
a
t
i
on
backbar
m
a
n
i
f
o
l
d
.
The
installation
of
non-interlock
v
a
po
r
i
ze
r
s
allows the
possible operation
of more
t
h
a
n
one
vaporizer
at the
same
t
i
m
e
.
5.3.4.1 Pre-use Check -
Interlock
S
ys
t
e
m
Check that the interlock mechanisms of all the vaporizers on
the manifold are working correctly, i.e. check that only one
vaporizer at a time can be turned on when the vaporizers are
adjacent to each other.
1
2