Prima 451 – User Manual 32
User Maintenance
6.3 Service
S
c
h
e
du
l
e
s
W
A
RN
I
N
G
1. Only fit components from preventive maintenance kits
supplied for the Prima 451 System.
2. If an individual device (e.g. ventilator, vaporizer,
absorber, AGSS receiver, suction controller) must
be removed for service or repair, make a note of the
serial number. When the device is refitted, check the
serial number against your recorded data.
6.3.1 Prima 451
Anaesthetic Ma
c
h
i
n
e
6 months Inspection and Function Check
12 months Inspection and Function Check. Replace
seals, etc. as required
2 years Fit components supplied in the Preventive
Maintenance Kit. Carry out an Inspection
and Function Check
Details of these service operations and ordering information
for spares are given in the Prima 451 Series Service Manual,
available only to engineers trained by the manufacturer.
6.3.2 Delta
V
a
po
r
i
ze
r
Refer to section 8 of the vaporizer user instruction manual.
Servicing includes a periodic leak test and calibration test,
and a major overhaul at 10 years (Halothane units: 5 years).
Test the vaporizers interlock system during the vaporizer
calibration test.
6.3.3
Monitoring
S
ys
t
e
m
s
Follow the service recommendations detailed in the monitor
user instruction manual.
6.3.4 Nuffield 200 or AV-S
V
e
n
t
il
a
t
o
r
6 months Inspection and Function Check
12 months Inspection and Function Check. Replace
seals, etc. as required.
Fit components supplied in the Preventive
Maintenance Kit (Nuffield 200)
2 years Fit components supplied in the Preventive
Maintenance Kit (AV-S MRI).
Inspection and Function Check
6.3.5 A200SP
Absorber
6 months Inspection and Function Check
12 months Inspection and Function Check. Replace
seals, etc. as required
2 years Fit components supplied in the Preventive
Maintenance Kit.
Inspection and Function Check
6.3.6
Anaesthetic
Gas
Scavenging
S
ys
t
e
m
Follow the service recommendations detailed in the AGSS
receiver user instruction manual.
6.3.7. SC760 Suction Controller
Follow the service recommendations detailed in section 6 in
the SC760 user instruction manual.