10
█
Other ancillary equipment
For instructions, refer to the respective manual provided with each equipment.
Category Description Model Name Brand Name
Irrigation Pump
EGA Series
EGA - 500P
PENTAX Medical
CO
²
Insufflator EGA - 501P
High Frequency Generator VIO Series
VIO 300D
ERBE
VIO 200S
Reprocessing before initial use/Reprocessing/Storage after use
█
Reprocessing before the initial use
Warning
The endoscope identified in this IFU is a reusable semi - critical device.
Since it endoscope and accessories are packaged non - sterile, they must be cleaned and high
level disinfected, or cleaned and sterilized according to the separate IFU (Reprocessing) of this
product before initial use as well as after each procedure and after repair. Insufficient reprocessing
may increase the risk of cross contamination.
Note
The wording "high - level disinfection" in this IFU defines the disinfection of the endoscope and the
accessories with a completely virucidal disinfectant.
█
Reprocessing
Warning
When using an endoscope and its accessories on patients with Creutzfeldt- Jakob disease (CJD)
or variant Creutzfeldt- Jakob disease (vCJD), use only dedicated instruments and equipment. The
instruments and equipment used on these patients must be discarded so that they can NOT be
used again on another patient. The pathogenic agents that cause this disease, which are called
“prions”, can NOT be destroyed or inactivated using the cleaning, disinfection, and sterilization
methods presented in this IFU. Please consult the guidelines that apply to your country or region
for more detailed information regarding the handling of prion- contaminated instruments.