Preparation and inspection
3
Inspection of the inlet seal (OF- B190)
Warning
NEVER use an inlet seal (OF - B190) that has any abnormality. Replace it with a new one.
Inlet seals are consumables. Using a damaged and /or worn inlet seal may result in lowered
suction function and potential reflux or dispersal of patient’s body fluids, posing a risk of
infection to the user.
1. Check the slit in the cap of the inlet seal (OF - B190) and the hole of the body of the inlet seal for
any abnormalities such as cracks, wear, chipping, and attachment or presence of foreign materials.
Check that the light does not shine through the slit of the cap.
Inlet Seal (OF - B190)
(3)
(1) Hole
(2) Slit
(3) Body
(4) Cap
Figure 3.21
2. Close the inlet seal as depicted in Figure 3.22.
Correct Incorrect
Figure 3.22
Note
PENTAX Medical PROfILE Single Use Endoscope Inlet Seal (OF - B215) can also be used. For
details on the inspection method, refer to the IFU provided with the inlet seal (OF - B215).