Preparation and inspection
3
Inspection of the irrigation function
Warning
Use sterile water for inspection of the irrigation function. Failure to do so may pose a risk of
infection.
1. Fill a syringe with sterile water.
2. Put the distal end of the endoscope to the clean container, and insert a syringe filled with sterile water
into the inlet seal (OF - B190) as shown in Figure 3.56.
Figure 3.56
3. Infuse sterile water into the instrument channel and check to ensure that it flows from the instrument
channel opening on the distal tip of the endoscope. Ensure that the effluent is free of foreign materials.
4. Remove the syringe from the inlet seal (OF- B190).
5. Fill the syringe with air and insert it into the inlet seal.
6. Flush the sterile water remaining inside the channel by pressing the syringe.
7. Remove the syringe from the inlet seal.