54
Preparation and inspection
3
Inspection of the instrument channel
Use a biopsy forceps for inspection of the instrument channel.
Prepare a biopsy forceps which has been cleaned and sterilized according to the manual provided with that
product and ensure to perform a pre - use inspection.
Warning
Do NOT use the endoscope if you feel a significant resistance when inserting a biopsy forceps.
It may result in damage to the inside of the channel and unforeseen events to patients and /or
medical professionals.
Caution
• Slowly and gently insert and withdraw the forceps from the inlet seal (OF- B190). Applying
strong force may cause endoscope damage.
• Keep the endoscope bending section as straight as possible when inserting the forceps, as it
may not be possible to pass the forceps through a highly angulated bending section.
1. Close the biopsy forceps cups by operating its handle.
CLOSE
CLOSE
OPEN
Figure 3.59
Note
Do not close the biopsy forceps cups tightly. Doing so may make its insertion into the instrument
channel difficult.