56
Preparation and inspection
3
4. Feed sterile water using the irrigation pump or syringe filled with sterile water attached to the luer
connector of the irrigation tube (OF - B113).
(1)
(2)
(3)
(1) Water Jet Connector Cap
(OF - B118)
(2) Irrigation Tube (OF - B113)
(3) Water Jet Check Valve Adapter
(OE - C12)
Figure 3.61
Note
When the irrigation tube (OF - B113) is connected with locking type of luer connector, ensure that
the luer connectors are properly locked. Do not use irrigation tube (OF - B113) if its luer connector
is damaged and /or if it is not properly connected.
5. Check that a certain amount of water flows out forward from the water jet nozzle at the distal end of
the endoscope. (It takes a few seconds until water comes out the first time.)
Figure 3.62
6. Check that there is no water leakage from the connection between the water jet port of the endoscope
and the water jet check valve adapter (OE- C12), or from the connection between the water jet check
valve adapter and the irrigation tube (OF - B113).
Figure 3.63