63
Directions for use
4
Water jet feeding
Warning
Use sterile water for water jet feeding. Failure to do so may pose a risk of infection.
Caution
Use minimum pressure for water feeding while observing the condition of the patient’s mucosa.
Water feeding with the excessive pressure may result in mucosal injury to the patient.
Use the irrigation pump by following its operation manual or by attaching a syringe to the luer connector of
the irrigation tube (OF - B113) and delivering water into it.
Note
When the irrigation tube (OF - B113) is connected with locking type of luer connector, ensure that
the luer connectors are properly locked. Do not use irrigation tube (OF - B113) if its luer connector
is damaged and /or if it is not properly connected.
Remote control
Caution
Do NOT apply strong force to the remote button from its side or in an oblique direction, as the
button may get stuck and become inoperable.
Operate the remote button for image capture, hardcopy, VCR recording, etc., as necessary.
Note
Leaving a finger on the remote button may result in unintentional pressing of the remote button,
causing it to operate.