EasyManuals Logo

PerkinElmer LAMBDA 265 Installation Instructions

PerkinElmer LAMBDA 265
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
09931228D
18
7. Select OK after checking the changed parameter values.
8. If the Peltier controller fails to communicate with the PC, change the COM Port Number in the
following steps.
9. Open Advanced Setting for COMx window following steps 1 to 6 in this procedure.
10. Select the COM Port number list to expand it and change the COM port number another one
which is not in use from COM 1 to COM 10.