a. Start Temp (°C): Shows the starting temperature of experiment.
b. Measurement Temp (°C): Shows the temperature at which the sample will be measured.
c. SP Temp (°C): Shows the current temperature while it goes up / down to the measurement
temperature.
d. Block Temp (°C): Shows the temperature of Cell Block.
e. Probe 1 Temp (°C): Shows the temperature of Probe 1.
f. Probe 2 Temp (°C): Shows the temperature of Probe 2.
g. Monitor: Monitored temperature is shown out of Block, Probe 1 and Probe 2.
h. Hold time (min): Shows the set holding time.
i. Elapsed time (m:s): Shows the elapsed time to reach the sampling temperature.
j. Status: Shows the status of experimental progress.
k. Error Range (°C): Shows the temperature tolerance between the sampling temperature and
monitored temperature. The measurement will be made only when the temperature tolerance
reaches within the set Error Range.
NOTE:
To pull out the cell easily, use the cell lifting knob.
Figure 17 Pulling out the cell
12. After the measurement is finished, the results are displayed in the result window. Save or print
results as required.
Troubleshooting
Power LED is not lit on
1. Check the connection of the power cable or the fuse. The fuse is located at the rear of the
instrument.
WARNING
There is risk of receiving a fatal electric shock if the fuses are replaced
with the power cord connected.
Il y a un risque d’électrocution si les fusibles sont remplacés ta
ndis que
le cordon d'alimentation est encore branché.

Other manuals for PerkinElmer LAMBDA 265

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer LAMBDA 265 and is the answer not in the manual?

PerkinElmer LAMBDA 265 Specifications

General IconGeneral
BrandPerkinElmer
ModelLAMBDA 265
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals