09931228D
9. After completing the parameter setup, select OK. Then the LED for On-Line is turned on, and it
will start heating up or cooling down to the set temperature in the Setup Single Cell Peltier.
10. Select Blank, and then the blank is measured when it reaches the set start temperature.
NOTE:
To monitor the probe temperature, the probes should be immersed in the sample, or to use the block
temperature for monitoring, the cell lid should be closed tightly.
11. Select Sample.
12. When the temperature reaches the target temperature, the measurement starts.
a. Start Temp (°C): Shows the starting temperature of the experiment.
b. Measurement Temp (°C): Shows the temperature at which the sample will be measured.