09931228D
6. Click on the setup button (pencil icon) in the Experiment Setup on the main screen of the
software.
7. Setup Single Cell Peltier window will be shown. Set parameters according to the experiment
conditions.
▶Temperature Control
a. Temperature (°C)
Enter the preset temperature for the experiment.
NOTE:
Starting the temperature of experiment needs to be set up in the method window.
b. Return to temperature (°C)
Enter the returning temperature after the experiment is finished.
c. Error Range (°C): Enter the temperature tolerance between the sampling temperature and
monitored temperature.The measurement will start when the temperature tolerance reaches
within the set Error Range.
Example: Set the starting temperature as 30°C and error range as 0.5°C then, the
monitoring temperature is set as the temperature of cell block. In this case, the
measurement will start when the temperature of cell block reaches between 29.5°C and
30.5°C.
NOTE:
When selecting probe for Monitor in Collect Temperature box, the slower the real heating speed may
be, the smaller the set Error range(°C) is.
▶Collect Temperature
a. Monitor
Select which temperature will be monitored in the display panel.