Flow SensorLED is lit on with an alarm sound
1. Check whether the tubing is bent or whether it is connected correctly, and then restart the Peltier
Controller.
2. Push the Air Vent Manualbutton on and check if coolant flows properly for about one minute
after the error occurs. If the Flow Sensor LED is continuouslyon with the alarm sounding, contact
PerkinElmer.
Pump LED is lit on with an alarm sound
1. Check whether the tubing is bent or whether it is connected correctly, and then restart the Peltier
Controller.
2. Push the Air Vent Manualbutton to be switched on and check if coolant flows properly for about
one minute after the error occurs. If the Pump LED is continuously on with the alarm sounding,
contact PerkinElmer.
1. Check whether the Interface connector is lined properly.
2. Check whether the MANUAL RUN/STOP button is pushed on. If yes, turn off the Manual
button.