44 Lambda 265, 365 and 465 ASX-260/280 AND ASX-520/560 Autosampler Users Guide
▶ Sample Rack Type
The following different types are available. Select the desired sample rack type.
1) 30 mm tubes (3 x 7 Positions)
2) 25 mm tubes (3 x 8 Positions)
3) 20 mm tubes (4 x 10 Position)
4) 15 mm tubes (5 x 12 Position)
5) 13 mm tubes (6 x 15 Position)
▶ COM Port No.: Click Port Find. The COM port no. will be selected automatically.
▶ Blank Position: Select the blank position.
▶ Sample Position
Samples are measured form Start Position to End Position in order.
- Start: Select the first sample vial for measurement.
- End: Select the last sample vial for measurement.
▶ Standard Position (For Quantitative Measurement): (1~10)
For Quantitative Measurement, Put the standards into the standard position to
make the calibration curve.
- Start: Select the first Standard Sample.
- End: Select the last Standard Sample.
▶ Probe Rinsing Time (sec)
If the Probe Rinsing Time is set, the probe will be rinsed after all measurements.
ex) the set time is '0', it won't be rinsed.
2) Peristaltic Pump Setting
▶ COM Port No.: Click Port Find. The COM port no. will be selected automatically.
▶ Pumping Speed (RPM): Enter the Pumping Speed. (0.1 ~100)
▶ Pumping Direction: Select the Pumping Direction. You can select either CW
(clockwise) or CCW (counter-clockwise).
▶ Pumping Time (sec): Enter the pumping time considering the length of tubing
from the auto sampler to the flow cell.