EasyManuals Logo

PerkinElmer Lambda 465 User Manual

PerkinElmer Lambda 465
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
56 . Lambda 265, 365 and 465 ASX-260/280 AND ASX-520/560 Autosampler Users Guide
COM Port No.: Click Port Find. The COM port no. will be selected automatically.
Blank Position: Select the blank position.
Sample Position
Samples are measured form Start Position to End Position in order.
- Start: Select the first sample vial for measurement.
- End: Select the last sample vial for measurement.
Standard Position (For Quantitative Measurement only): (1~10)
For Quantitative Measurement, Put the standards into the standard position to
make the calibration curve. It is not available in Scan and Wavelength
Program mode.
- Start: Select the first Standard Sample.
- End: Select the last Standard Sample.
Probe Rinsing Time (sec)
If the Probe Rinsing Time is set, the probe will be rinsed after all measurements.
ex) the set time is '0', it won't be rinsed.
2) Peristaltic Pump Setting
COM Port No.: Click Port Find. The COM port no. will be selected automatically.
Pump Model: Pump model is fixed. (Version1).
Pumping Speed(RPM): Enter the Pumping Speed. (0.1~100)
Pumping Direction: Select the Pumping Direction. You can select either CW
(clockwise) or CCW (counter-clockwise).
Pumping Time (sec): Enter the pumping time considering the length of tubing
from the auto sampler to the flow cell.
Stabilizing Time (sec)
To make stabilization after the blank or sample solution is filled up the cell, enter
the suitable stabilizing time. It recommends more than 20 sec. If the time is set as
'0', it will be scanned directly without the stabilization.

Other manuals for PerkinElmer Lambda 465

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer Lambda 465 and is the answer not in the manual?

PerkinElmer Lambda 465 Specifications

General IconGeneral
BrandPerkinElmer
ModelLambda 465
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals