Installation and storage
This section contains information regarding the installation and storage of the TriNEST Microplate incubator and shaker
instrument.
Unpacking
Unpack the instrument carefully and check for any damage that may have occurred during transportation. If any parts
are damaged or missing, contact your PerkinElmer representative.
Note: Pictures in this manual are for illustration purposes only and may not be an exact representation of the product.
The actual product appearance may vary depending on model and country of delivery.
Environmental requirements
The instrument is intended to be used in a clean and controlled laboratory environment.
The instrument should be positioned to avoid exposure to excess dust, vibrations, strong magnetic fields, direct
sunlight, draft, excessive moisture or large temperature fluctuations.
The laboratory table must be strong and have a flat and non‑slip surface.
Leave sufficient clearance (10 cm) around the instrument for adequate air circulation.
Ambient temperature range during operation: 15 - 35°C
Relative humidity during operation: 10 - 75%
Installation
Use the power cable supplied to connect the instrument to the mains supply. This supply must have a protective earth.
The appliance connector socket is at the back of the instrument.
To connect the instrument to a different mains system, you must use an authorized grounded adapter. Note! The power
supply between the mains and the instrument works with voltages between 100V and 240V.
Warning: Maximum three stacked instruments. The attachment parts (Connection Plate 2300073600) supplied with
the instrument are required when stacking instruments on top of each other. Instrument stacking installation is only
to be performed by personnel properly trained and authorized by PerkinElmer.
Figure 2: Two TriNEST instruments stacked on top of each other, front view
TriNEST Microplate incubator and shaker
Installation and storage
7