Display
The display shows the following symbols (from top to bottom):
Program number (01 to 20) and Program
name (up to 6 characters)
Target incubation temperature in the
range of 20 to 70°C.
An arrow pointing up shows the target has
not yet been reached, but the temperature
is rising towards it. An arrow pointing down
shows the temperature is above the target
and is cooling down towards it.
Incubation time -in hours and minutes
(max. 48 hours).
Shaking speed -400 to 1200 rpm in steps
of 100; or OFF (no shaking)
Interval shake time -in minutes and
seconds (00:00 = no shaking)
Interval rest time - in minutes and
seconds (00:00 = no pause)
Note: During operation, an arrowhead
appears next to functions that are
occurring.
Near the bottom of the display you can see the status of the three plate holders:
TriNEST Microplate incubator and shaker
Functional description
11