Owner´s Manual Permobil C500 R-net Diagnostics
83
3.1 Joystick Error
The most common cause of this trip is if the joystick is deflected away from cen-
ter before and during the time the control system is switched on. The joystick
displaced screen will be displayed for 5 seconds, if the joystick is not released
within that time then a trip is registered. Although a trip screen is not displayed
the system log will show the trip and numbers of occurrences.
• Ensure that the joystick is centered and power-up the control system.
If the trip is still present then the joystick or Joystick Module may be defective.
Refer to Section 5.
3.2 Low battery
This occurs when the control system detects that the battery voltage has fallen
below 16V.
• Check the condition of the batteries and the connections to the control
system.
If the trip is still present after the batteries and connections have been checked,
then the Power Module may be defective. Refer to section 5.
3.3 High Battery Voltage
This occurs when the control system detects that the battery voltage has risen
above 35V. The most common reasons for this are overcharging of the battery
or bad connections between the control system and the batteries.
• Check the condition of the batteries and the connections to the control system.
If the trip is still present after the batteries and connections have been checked,
then the Power Module may be defective. Refer to Section 5.