EasyManua.ls Logo

Permobil C500 - Page 90

Permobil C500
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Owner´s Manual Permobil C500 R-net Diagnostics
90
4. Basic Test
After a repair has been completed, the following tests should be carried out.
These are minimum recommendations, depending on the nature of the original
trip then additional tests may be required.
4.1 General Inspection
Make sure all connectors are securely mated.
Check the condition of all cables and connectors for damage.
Check the rubber gaiter or boot around the base of the joystick shaft for
damage. Check visually only, do not handle the gaiter.
Make sure that all components of the control system are securely mounted.
Do not overtighten any securing screws.
WARNING
These tests are a minimum recommendation only. It is the responsibility of the service person(s)
to perform other tests relevant to the original trip and wheelchair type are carried out. Refer to the
wheelchair’s Service Manual for exact information of other tests. Permobil accept no liability for
losses of any kind arising from the carrying out of the described tests, or from not carrying out
additional relevant tests.
WARNING
These tests should be conducted in an open space and a restraining device such as a seat belt
should always be used. Permobil accepts no liability for losses of any kind arising from failure to
comply with this condition.