MCO-170AICD-PE
MCO-170AICUVD-PE
Please read the operating instructions carefully before using this product, and keep the operating
instructions for future use.
See page 101 for model number.
Operating Instructions
CO
2
Incubator

Table of Contents

Questions and Answers

  • J
    Jeffrey ThompsonAug 20, 2025
    How to troubleshoot a Phcbi Accessories incubator that does not operate at all?
    • M
      mary95Aug 20, 2025
      If your Phcbi Accessories incubator isn't operating, start by checking these potential issues: * Ensure the incubator is properly plugged into a functioning power outlet. * Verify there isn't a power failure or a tripped circuit breaker. * Confirm that the removable power supply cord is securely connected to the port on the lower right side of the cabinet.
  • S
    Sharon PattersonAug 22, 2025
    What to do if the CO2 gas density does not match the set value in my Phcbi Accessories?
    • B
      bryan82Aug 22, 2025
      If the CO2 gas density in your Phcbi Accessories incubator doesn't match the set value, check the following: * Ensure the secondary pressure for the gas regulator is at the specified value of 0.03 MPa(G)~0.1 MPa(G) (0.3 kgf/cm2(G)~1 kgf/cm2(G), 4.4 psi(G)~14.5 psi(G)). * Check for any blockages in the gas tube. * Make sure the duct is securely attached. * Confirm that the fan is properly attached. * Verify that the incubator is not operating near appliances that generate electromagnetic waves.
  • C
    Courtney JohnsonAug 26, 2025
    What causes the chamber temperature to not match the set value in my Phcbi Accessories incubator?
    • H
      Hector GarciaAug 26, 2025
      If the chamber temperature of your Phcbi Accessories incubator does not match the set value, consider these possible causes: * Ensure the ambient temperature is no more than 5 °C different from the set chamber temperature. * Verify that the outer door is closed and the inner door is also closed. * Check that the incubator is not operating near equipment that generates electromagnetic waves.
  • D
    Donald PerezAug 27, 2025
    Why is the CO2 gas density slow to recover in my Phcbi Accessories incubator?
    • A
      Aaron MeyerAug 28, 2025
      If the CO2 gas density in your Phcbi Accessories incubator is slow to recover, consider the following: * The incubator CO2 gas piping uses a HEPA filter. * Check if there is little gas remaining in the CO2 gas cylinder. * Verify the secondary pressure for the gas regulator is at the specified value of 0.03 MPa(G)~0.1 MPa(G) (0.3 kgf/cm2(G)~1 kgf/cm2(G), 4.4 psi(G)~14.5 psi(G)). * Ensure the gas tube is not blocked. * Confirm the duct is securely attached and the fan is properly attached.
  • J
    jonathanhooperAug 30, 2025
    Why is a large quantity of CO2 gas being consumed by my Phcbi Accessories incubator?
    • J
      jamessheltonAug 30, 2025
      If your Phcbi Accessories incubator is consuming a large quantity of CO2 gas, investigate the following: * Check if the outer and inner doors are being opened and closed frequently. * Inspect connectors for gas leaks due to deterioration of the gas tube or pinhole leaks. * Examine the packing seal for the inner door for defects. * Ensure the access hole is closed.
  • R
    Rachel GambleSep 1, 2025
    What to do if the chamber humidity does not rise in my Phcbi Accessories incubator?
    • J
      John EvansSep 1, 2025
      If the chamber humidity in your Phcbi Accessories incubator is not rising, ensure there is enough sterile distilled water in the humidifying pan.
  • C
    campbelltheresaSep 4, 2025
    What could cause normal cultures to be impossible with a suspect CO2 gas density in my Phcbi Accessories?
    • A
      Anita RileySep 4, 2025
      If normal cultures are not possible and the CO2 gas density is suspect in your Phcbi Accessories incubator, verify the following: * Check if the ambient air environment around the incubator is normal. * Ensure there is no source of polluted gas in the vicinity.
  • L
    Linda PhillipsSep 7, 2025
    Why does the outer door of my Phcbi Accessories incubator not open?
    • J
      Julie DixonSep 7, 2025
      The outer door of your Phcbi Accessories incubator may not open because: * When the power switch is OFF, the electric lock is engaged. * During dry heat sterilization, the outer door is electrically locked.
  • M
    markriddleSep 9, 2025
    Why is CO2 gas not being injected into my Phcbi Accessories incubator?
    • P
      paul15Sep 9, 2025
      CO2 gas may not be injected into your Phcbi Accessories incubator because: * The CO2 control method is the ON-OFF method. * Gas injection is delayed until the CO2 sensor's temperature stabilizes.
  • M
    Monica HallSep 12, 2025
    Why does dry heat sterilisation end in the middle with an error in my Phcbi Accessories?
    • H
      hillrachelSep 12, 2025
      Dry heat sterilization might end prematurely in your Phcbi Accessories incubator due to: * Missing silicon caps on the access port. * The inner door being open.

Summary

Introduction and Safety Precautions

Intended Use and Precautions

Defines the product's purpose and general usage precautions.

Precautions for Safe Operation

Critical safety advice to prevent injury or damage.

Incubator Components

Unit Components Overview

Detailed components of the incubator unit.

LCD Touch Panel Interface

Explanation of the incubator's touch screen interface.

Remote Alarm Terminal

Function and connection of the remote alarm terminal.

Installation Guidelines

Installation Site Selection

Guidelines for selecting a suitable location for the incubator.

Connecting CO2 Gas Cylinder

Procedure for connecting the CO2 gas cylinder.

Before Commencing Operation

Initial Cleaning Method

Steps for cleaning the incubator before initial use.

Removing Inner Attachments

Instructions for safely removing internal components.

Installing Inner Attachments

Procedures for correctly re-installing internal components.

Filling Humidifying Pan

Guide on how to fill and maintain the humidifying pan.

For Better Cultivation

Precautions for Cultures

Essential precautions to ensure successful cell cultures.

Preventing Contamination

Methods to avoid and prevent contamination within the chamber.

Correct Operation

Basic Operation on LCD Touch Panel

Basic Parameters

Numerical Input to Input Window

How to input numerical values using the interface.

Setting Temperature, CO2, and High Limit Alarm

Configuring essential incubator parameters like temperature and CO2.

Setting Key Lock

How to enable and configure the key lock function.

Removing Key Lock

Procedure to disable or remove the key lock.

Alarm Parameters

Alarm Setting #1 Configuration

Setting up temperature and CO2 density alarms.

Alarm Setting #2 Configuration

Configuring alarm delays, ring back, and remote alarms.

Operation;Alarm;Sterilisation Log

Setting Log Interval

Adjusting the frequency of data logging.

Displaying Operation Log

How to view logged operational data graphically.

Exporting Operation Log

Steps to export operational data to external storage.

Displaying Alarm Log

Viewing logged alarm events.

Exporting Alarm Log

Exporting alarm log data.

Displaying Sterilisation Log

Viewing logged sterilization cycles.

Exporting Sterilisation Log

Exporting sterilization log data.

Other Parameters

Setting Date and Time

Adjusting the incubator's date and time settings.

Setting Brightness and Sleep

Configuring display brightness and sleep mode.

Dry Heat Sterilisation

Dry Heat Sterilisation Procedure

Procedure for performing dry heat sterilization.

Electric Lock Functions

Setting User-ID

Registering User-IDs for access control.

Setting Auto Lock

Configuring the automatic door locking function.

Removing Auto Lock

Disabling the automatic door locking.

Using Unlock Key

How to use the unlock key, especially during power failure.

UV Lamp Parameters

Using UV Lamp

Guide to operating the UV lamp for disinfection.

Setting UV Lamp ON Period

Configuring the duration the UV lamp stays on.

Lighting UV Lamp for 24 Hours

Procedure for continuous UV lamp operation.

Gas Auto Changer (Option)

Connecting CO2 Gas Cylinder

Connecting CO2 gas cylinders for the auto changer.

Automatic CO2 Gas Supply Line Changeover

How the system automatically switches gas lines.

Manual CO2 Gas Supply Line A;B Changeover

Manually switching between gas supply lines.

STD Gas Auto Calibration Kit (Option)

STD Gas Setting

Setting the standard CO2 density for calibration.

STD Gas Calibration

Performing CO2 density calibration.

Alarms, Safety, and Self-Diagnosis

Incubator Alarms and Safety Functions

Overview of incubator alarms and safety features for cultivation.

Dry Heat Sterilisation Safety

Safety functions and alarms during sterilization.

Troubleshooting

Disposal of Unit

Phcbi MCO-170AICD-PE Specifications

General IconGeneral
BrandPhcbi
ModelMCO-170AICD-PE
CategoryAccessories
LanguageEnglish