MCO-170AICDL
MCO-170AICUVDL
Please read the operating instructions carefully before using this product, and keep the operating
instructions for future use.
See page 101 for model numbers.
Operating Instructions
CO
2
Incubator
This product is for research use only.
Not for clinical diagnosis or treatment of humans or animals.

Table of Contents

Questions and Answers

  • K
    Kenneth LawsonAug 20, 2025
    Why is my Phcbi Laboratory Equipment incubator not operating at all?
    • L
      Linda MasseyAug 20, 2025
      If your Phcbi Laboratory Equipment incubator isn't operating, first ensure it's properly plugged in and that there is no power failure or tripped circuit breaker. Also, check that the removable power supply cord is correctly connected to the port on the lower right side of the cabinet.
  • P
    pwaltersAug 23, 2025
    What to do if the CO2 gas density does not match the set value in Phcbi Laboratory Equipment?
    • D
      Derek HowardAug 23, 2025
      If the CO2 gas density in your Phcbi Laboratory Equipment does not match the set value, check the following: * Ensure the secondary pressure for the gas regulator is at the specified value of 0.03 MPa(G)~0.1 MPa(G) (0.3 kgf/cm2(G)~1 kgf/cm2(G), 4.4 psi(G)~14.5 psi(G)). * Verify that the gas tube is not blocked. * Confirm that the duct is securely attached to the 4 points hooks. * Ensure the fan is properly attached and pushed all the way to the motor shaft. * Also, consider whether the incubator is operating near equipment that generates electromagnetic waves.
  • M
    Molly GibsonAug 26, 2025
    What to do if the chamber temperature does not match the set value in Phcbi Laboratory Equipment?
    • M
      Michael LewisAug 26, 2025
      If the chamber temperature in your Phcbi Laboratory Equipment does not match the set value, verify that the ambient temperature is not less than 5 °C different from the set chamber temperature. Also, ensure that the outer door is closed and the inner door is not left open. Check if the incubator is operating beside equipment that generates electromagnetic waves.
  • F
    fred81Aug 28, 2025
    Why is CO2 gas not being injected into my Phcbi Laboratory Equipment?
    • J
      jennifermcintoshAug 28, 2025
      The CO2 control method for the Phcbi Laboratory Equipment incubator is the ON-OFF method. CO2 gas is injected intermittently as the gas density in the chamber approaches the set value. Injections may be stopped for periods of approximately 15 seconds, but that is not an error. Also, the gas is not injected until the temperature of the CO2 sensor becomes stable enough (approximately 1 hour) after turning ON the power switch or recovering from power failure.
  • A
    Aimee SimmonsAug 31, 2025
    Why is a large quantity of CO2 gas being consumed by my Phcbi Laboratory Equipment?
    • D
      douglas23Aug 31, 2025
      A large quantity of CO2 gas being consumed by your Phcbi Laboratory Equipment may be due to frequently opening and closing the outer and inner doors. Check for gas leaks from connectors due to deterioration or pinhole leaks in the gas tube; consider replacing the gas tube annually. Also, inspect the packing seal for the inner door and ensure the access hole is closed.
  • A
    anna77Sep 2, 2025
    What to do if data cannot be exported to the USB flash drive from Phcbi Laboratory Equipment?
    • C
      Christina DaySep 3, 2025
      If you cannot export data to the USB flash drive from your Phcbi Laboratory Equipment, please check the following: Ensure the USB flash drive is inserted properly. Verify that data exists for the specified time period. Confirm the USB flash drive is not full and is formatted in FAT16 or FAT32 format. Do not use a USB flash drive that requires a password or has a capacity of more than 32 GB.
  • B
    Brenda FreemanSep 5, 2025
    What to do if dry heat sterilisation ends in the middle by an error in Phcbi Laboratory Equipment?
    • J
      jeffreykingSep 5, 2025
      If dry heat sterilisation ends prematurely in your Phcbi Laboratory Equipment, make sure that silicon caps are attached to the access port and that the inner door is closed.
  • A
    Ann StrongSep 8, 2025
    Why won't the outer door open on my Phcbi Laboratory Equipment?
    • D
      Dr. Nicholas Woods DVMSep 8, 2025
      If the outer door of your Phcbi Laboratory Equipment won't open, it may be because the electric lock is engaged when the power switch is OFF. Turn the power switch ON or use the accessory key to unlock the electric lock. Also, during dry heat sterilization, the outer door is electrically locked and cannot be opened.
  • F
    francis47Sep 12, 2025
    Why doesn't the chamber humidity rise in my Phcbi Laboratory Equipment?
    • J
      john41Sep 12, 2025
      If the chamber humidity does not rise in your Phcbi Laboratory Equipment, ensure there is enough sterile distilled water in the humidifying pan.
  • C
    Carly HernandezSep 13, 2025
    Why does the UV lamp light when the outer door is open on my Phcbi Laboratory Equipment?
    • L
      Louis GibsonSep 14, 2025
      If the UV lamp lights when the outer door is open on your Phcbi Laboratory Equipment, check if something is pushing the door switch.

Phcbi MCO-170AICDL Specifications

General IconGeneral
BrandPhcbi
ModelMCO-170AICDL
CategoryAccessories
LanguageEnglish