MCO-5MSeries
Please read the operating instructions carefully before using this product, and keep the operating
instructions for future use.
See page 52 for model number.
Operating Instructions
O
2
/CO
2
Incubator

Questions and Answers

  • N
    Nancy FordAug 13, 2025
    How to troubleshoot a Phcbi Accessories unit that won't turn on?
    • W
      wagnermarkAug 13, 2025
      If your Phcbi Accessories unit isn't operating, make sure it's correctly plugged into the power outlet. Also, verify that the circuit breaker at the power source is active and that there isn't a power failure. Finally, check if the power supply cord is properly connected to the port on the rear of the cabinet.
  • C
    Carrie RobersonAug 17, 2025
    Why is gas consumption so high in my Phcbi MCO-5M-PE unit?
    • L
      Larry FisherAug 17, 2025
      If the gas consumption of your Phcbi Accessories unit is too high, it could be due to several reasons. First, check if the door is being opened too frequently. Second, inspect the gas tube connections for any leaks or pinholes, and consider replacing the gas tube annually. Third, ensure the inner door gasket is completely sealed. Lastly, verify that the access port at the upper left corner is closed.
  • R
    Rodney WeaverAug 25, 2025
    Why doesn't the gas density match the set value in my Phcbi Accessories?
    • A
      Angela AllenAug 25, 2025
      If the gas density in your Phcbi Accessories unit does not match the set value, it might be because the secondary pressure isn't set correctly. Another possible reason is that the gas tube is clogged or kinked. Ensure that the humidifying pan is filled with sterile distilled water, as using other types of water can affect gas density.
  • K
    Karen Reyes DDSAug 28, 2025
    What to do if chamber temperature is not equal to the set temperature in Phcbi MCO-5M-PE?
    • M
      Matthew PattersonAug 28, 2025
      If the chamber temperature of your Phcbi Accessories unit is not equal to the set temperature, ensure that the ambient temperature is always at least 5°C less than the set temperature. Also, check that the outer door was not closed while the inner door was left open.
  • J
    Jeffery MontesSep 1, 2025
    Why is the chamber humidity not rising in my Phcbi Accessories?
    • A
      Amy Saunders DDSSep 1, 2025
      If the chamber humidity in your Phcbi Accessories unit isn't rising, make sure the humidifying pan is filled with sterile distilled water. Always use sterile distilled water to ensure proper humidity levels.
  • A
    Andrea TorresSep 5, 2025
    Why normal cultivation cannot be done and chamber gas density is suspect in Phcbi Accessories?
    • S
      Steven CooperSep 5, 2025
      If normal cultivation cannot be done and chamber gas density is suspect in your Phcbi Accessories unit, it might be due to an abnormal environment around the unit or because the unit is installed in an enclosed space.
  • D
    Dennis SmithSep 8, 2025
    How to unlock the keys on a Phcbi MCO-5M-PE unit?
    • M
      Mr. Robert WallSep 9, 2025
      If the key operation is disabled on your Phcbi Accessories unit, the key lock function is likely set to the ON mode. Disable the key lock function to restore key operation.
  • C
    Cynthia SalinasNov 20, 2025
    What to do if the alarm function operates on my Phcbi MCO-5M-PE unit?
    • D
      diana53Nov 20, 2025
      If the alarm function is operating on your Phcbi Accessories unit, it could be due to several reasons: * The chamber temperature isn't matching the set value. * The chamber CO₂/O₂ density doesn't match the set value. * The secondary pressure of the gas regulator isn't at the set value. * The gas tube isn't securely connected between the gas regulator and the unit. * The high limit alarm temperature isn't set at least 1°C higher than the set chamber temperature. * The set temperature value was changed, or the door was left open for too long. * The gas tube has slipped off, or there's a gas leak. * The set value of the gas density was changed. * The gas cylinder is empty; check the primary pressure weekly, and replace the cylinder if the CO₂ primary pressure is less than 3...

Summary

Safety Precautions for Incubator Operation

Incubator Component Identification

Incubator Control Panel and Keypad

Remote Alarm Terminal Functionality

Incubator Installation Site Requirements

Incubator Installation Procedures

Connecting CO2 Gas Supply

Step-by-step guide for connecting the CO2 gas cylinder, including safety warnings and pressure settings.

Connecting N2 or O2 Gas Supply

Procedure for connecting N2 or O2 gas cylinders, specifying gas type, pressure, and safety considerations.

Precautions for Cell Culture Procedures

Initial Incubator Cleaning and Preparation

Routine Incubator Maintenance Schedule

Initial Startup and Correct Operation

Control Panel Key Operations and Settings

Automatic Gas Cylinder Changeover System

UV Lamp System Operation and Maintenance

Alarms, Safety Features, and Self-Diagnosis

Incubator Calibration Procedures

Calibrating Incubator Temperature

Step-by-step guide for calibrating the chamber temperature for accurate readings.

Calibrating O2 Density Levels

Procedure for performing zero and span calibration for the O2 density sensor.

Incubator Troubleshooting Guide

Service Safety Check Sheet

Phcbi MCO-5M-PE Specifications

General IconGeneral
CategoryAccessories
ModelMCO-5M-PE
ManufacturerPHCbi
TypeCO2 Incubator
CO2 Range0 to 20%
CO2 Control Accuracy±0.15%
Temperature Range5°C above ambient to 50°C
Temperature Uniformity±0.3°C