Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic
compatibility (EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying tables.
Only use this Philips Avent single/double electric breast pump Plus in combination with the adapter
supplied.
Emissions Compliance Electromagnetic environment –
guidance
RF emissions
CISPR 11
Group 1 The device does not use RF energy for its
intended function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Complies
Class A Category
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
The device is suitable for use in locations in
residential environments and in establishments
directly connected to a low voltage power
supply network which supplies buildings used
for domestic purposes.
Guidance and manufacturer's declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Plus is intended for use in the electromagnetic
environment listed below. The user of the device should ensure that it is used in such an environment to
prevent improper operation.
Immunity test Immunity test level Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Floors can be a source of static electricity. When
the air is very dry (<20% RH) the user is advised
to prefer to use the device in a room with
concrete or wooden floors (instead of using it in
a room with a synthetic floor).
Radiated RF EM fields
IEC 61000-4-3
10 V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
385 (27), 450 (28), 710 (9),
745 (9), 780 (9), 810 (28),
870 (28), 930 (28), 1720
(28), 1845 (28), 1970 (28),
2450 (28), 5240 (9), 5500
(9), 5785 (9) MHz (V/m)
The device is immune against any RF wireless
communication device such as smart phone, Wi-
Fi or Bluetooth device at a distance of 30cm or
more. Avoid a closer distance as this may result
in improper operation.
Rated power frequency
magnetic fields IEC
61000-4-8
30 A/m
50 Hz or 60 Hz
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV
100 kHz repetition
frequency
Mains power quality expected of a typical
commercial or residential environment.

Table of Contents