Single electric breast pump Double electric breast pump
Vacuum level
Stimulation: -60 to -200 mbar (-45 to -150 mmHg)
Expression: -60 to -360 mbar (-45 to -270 mmHg)
Adjustable in steps of 20 mbar
Intermittent vacuum
Cycle speed
Stimulation: 105-120 cycles/min, depending on vacuum level
Expression: 53-85 cycles/min, depending on vacuum level
Motor unit rated input 5 V direct current / 1.1 A 5 V direct current / 1.8 A
Mains adapter rated input
External mains adapter 100-240 V alternating current
50/60 Hz
400 mA
Mains adapter rated output 5 V direct current / 1.1 A 5 V direct current / 1.8 A
Adapter type number
S009AHz050yyyy The letters "yyyy" represent the output current from
0100 (1000mA) to 0180 (1800mA), increments in steps of 100mA. The
letter "z" denotes type of plug, it can be: U, V, B, S, C, A, D, K, T, I, E, F.
Power port
Micro-USB
Operating conditions
Temperature: 5°C to 40°C
Relative humidity: 15 to 90% (non-condensing)
Storage and transportation
conditions
Temperature: -25°C to 70°C
Relative humidity: up to 90% (non-condensing)
Operating and storage
pressure
700-1060 hPa (<3000m altitude)
Net weight 230g 310g
External dimensions
Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
Device classification
Applicable classifications:
Degree of protection against electric shock;
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specified as medical electrical system
according to IEC 60601-1.
Materials
ABS, silicone (motor unit)
Polypropylene, silicone (other parts)
Service life 2 years
Operational life 500 hours
Mode of operation Continuous operation
Ingress protection IP22 (motor unit), IP20 (adapter)