Chapter A: Installing Multiple ITSs at a Single Hospital Site
Sync Network Requirements for Multiple ITSs
Note the design rules for synchronization of multiple IntelliVue Telemetry Systems:
A common Sync Network is required if the ITSs are on adjacent floors, or if the
linear distance between system access points on the same floor is less than
300 feet.
The ITSs do not have to be synchronized to each other if they are not on
adjacent floors, or if the linear distance between system access points on the
same floor is greater than 300 feet.
Figure A-3: Common Sync Network Required
Figure A-4: Common Sync Network Not Required