Chapter 1: Overview
Note the following guidelines for installing the ITS within a routed ICN topology:
An ITS subnet can be connected up to 22 ICN Database Domains via routers.
Up to 320 ITS Standard Access Points may be installed on a routed ICN
topology.
Up to 320 ITS Core Access Points and up to 640 Remote Antennas (i.e., two per
Core AP) may be installed on a routed ICN topology.
ITS Standard and Core APs may exist together on a routed ICN topology so long
as the maximum number of APs does not exceed 320.
Only the 1.4 GHz ITS may be installed within a routed ICN topology. The 2.4 GHz
ITS does not support installation within a routed ICN topology.
Refer to Table 1-10 for a list of the device IP address assignments used in a routed
ICN configuration where the ITS infrastructure is installed as a separate subnet to
which up to 22 Database Domains have access via routers.
Note the following regarding Table 1-10:
“n” represents the network number and starts at 0 for single ICN systems (i.e.,
Database Domains). This variable increments by 8 from there for additional
IntelliVue Clinical Networks. For example, for ICN 2, “n” would equal 8, for ICN 3,
“n” would equal 16, and so on.
Route statements are generated (in instances with a Router and without) at the
completion of the Config Wizard.