2-30
Chapter 2: Planning Your ITS Deployment
APC Configuration Worksheet for Non-routed ICN Topology
BOOTP/DHCP Server Configuration
APC Static Address Configuration
APC Name MAC Address IP Address Subnet Mask Default Gateway
Range:
172. 31.(n+1).0 -
172. 31.(n+1).63
255.255.248.0 172.31.(n + 3).0
(DBS)
255.255.248.0
255.255.248.0
255.255.248.0
255.255.248.0
255.255.248.0
255.255.248.0
255.255.248.0
255.255.248.0
255.255.248.0
NON-ROUTED: Range 1 - Transceivers and Wireless Bedsides
Setting Rules Factory Default Site Modification
MAC Address Base 00:09:fb:06:00:00 00:09:fb:06:00:00 00:09:fb:06:00:00
MAC Address Mask ff:ff:ff:0f:00:00 ff:ff:ff:0f:00:00 ff:ff:ff:0f:00:00
IP Address Range (Minimum) 172.31.(n + 6).0 172.31.6.0
IP Address Range (Maximum) 172.31.(n + 6).255 172.31.6.255
Subnet Mask 255.255.248.0 255.255.248.0 255.255.248.0
Default Gateway 172.31.(n + 3).0 (DBS) 172.31.3.0
DNS Server IP Address 0.0.0.0 0.0.0.0 0.0.0.0
NON-ROUTED: Range 2 - Access Points
Setting Rules Factory Default Site Modification
MAC Address Base 00:09:fb:05:00:00 00:09:fb:05:00:00 00:09:fb:05:00:00
MAC Address Mask ff:ff:ff:0f:00:00 ff:ff:ff:0f:00:00 ff:ff:ff:0f:00:00
IP Address Range (Minimum) 172.31.(n + 2).128 172.31.2.128
IP Address Range (Maximum) 172.31.(n + 2).255 172.31.2.255
Subnet Mask 255.255.248.0 255.255.248.0 255.255.248.0
Default Gateway 172.31.(n + 3).0 (DBS) 172.31.3.0
DNS Server IP Address 0.0.0.0 0.0.0.0 0.0.0.0