Environmental Protection
6 Chapter 2 - Routine Maintenance
The battery can be recharged while the monitor is in use, in
which case, the battery requires 14 hours to be recharged.
The battery can require a full discharge/charge cycle to restore
normal capacity.
If the monitor operates
1
for less than one hour on battery power before the
low battery alarm occurs, the battery should be conditioned.
If the same symptom persists even after the battery is conditioned and
indicating a full charge, the battery should be replaced.
Refer to Chapter 6, “Disassembly Guide”.
Environmental Protection
Follow local governing ordinances and recycling plans regarding disposal or
recycling batteries and other device components.
1. See “Battery Performance Test” on page 14 for typical battery operating times
and conditions.