Equipment Needed
10 Performance Verification
Equipment Needed
The following table lists the equipment required for performance verification.
Table 3 Required Test Equipment
Required For
Equipment Description A1 A3
Digital multimeter
(DMM)
Fluke Model 87 or equivalent
··
Sensor extension cable M4787A
··
Finger clip sensor M4789A
··
Oxisensor® II adhesive
sensor
D-25
··
ECG cable M3913A M3923A and
M3924A only
·
ECG electrodes Standard M3923A and
M3924A only
·
ECG leads M3914A (IEC) or M3915A
(AAMI)
M3923A and
M3924A only
·
NiBP tubing M3918A
··
NiBP cuff 40401C
··
Pulse oximeter tester Nellcor Puritan Bennett SRC-2
··
ECG simulator Dynatech Nevada medSim
300 or equivalent
··
NiBP simulator Bio-Tek BP Pump or
equivalent
··
Temperature simulator medSim 300 or equivalent M3922A and
M3924A only
·
Respiration simulator medSim 300 or equivalent
Safety analyzer Bio-Tek 601 Pro or equivalent
··
Stopwatch Manual or electronic
··