Disassembly Procedures (A1)
66 Chapter 6 - Disassembly Guide
compartment, and the battery cover.
Separation of Front and Rear Case Assemblies
This section describes the procedures required to separate the front and rear
case assemblies of the A1 monitor.
Procedure to separate the front and rear case assemblies:
1. Remove the SpO
2
connector hood by squeezing the sides to release the
detents holding the hood in place.
During reassembly, the notch on the SpO
2
connector hood should be
facing down.
2. Use a Phillips-head screwdriver to remove the four screws fastening
together the front and rear case assemblies.
3. Separate the main front and rear case assemblies.
4. If the rear cover gasket seal is to be replaced, remove it.
Foot
cushions
Battery
cover
SpO
2
connector
hood