Disassembly Procedures (A1)
Chapter 6 - Disassembly Guide 65
2. Remove the temperature module housing and the battery cover.
3. The probe sensor contact switch is mounted on the outer side of the
battery cover. As required, disconnect the leads from the probe sensor
switch to the rear panel connector.
4. Remove the switch.
5. Disconnect the spade terminal connectors from the battery terminals.
Caution During reassembly, pay particular attention to the polarity
of the battery cable before reattaching. The polarity is
marked on the circuit board. Red connects to the +ve
terminal, black connects to the –ve terminal. If battery cable
polarity is reversed, it is likely that circuit damage will
occur.
6. Remove the battery.
7. As required, remove the battery cushions on the inside of the battery
Removed
temperature
module
housing
Battery
cover
Temperature
probe grommet
Spare probe
sensor switch
Probe sensor
switch