Disassembly Procedures (A1)
76 Chapter 6 - Disassembly Guide
Procedure to separate the patient connector PCB from the
connector panel:
1. Use a Phillips-head screwdriver to remove four screws fastening the two
assemblies together.
Two of the screws are accessible on the face of the connector PCB, two
are accessible through access holes in the PCB.
2. Separate the two assemblies.
Patient connector PCB assembly
Foot of moulded connector panel after removal of Tinnerman nut
Main PCB
assembly
Connector
panel
Patient
connector
PCB