Hardware and Software Tests
14 Performance Verification
Hardware and Software Tests
Hardware and software testing includes the following tests applicable to the
indicated models in the series:
SpO
2
Testing
SpO
2
testing includes the following tests:
n Alarms and Alarm Silence
n Heart Rate Tone Volume Control
n Dynamic Operating Range
n LED Excitation Test
Alarms and Alarm
Silence
1. Connect the SRC-2 pulse oximeter tester to the M4787A sensor cable and
connect the cable to the monitor.
2. Set SRC-2 as follows:
3. Press the ON/STANDBY button to turn the monitor on.
4. After the normal power-up sequence, verify that the SpO
2
% display initially
Test Applies to Model(s)
SpO
2
Testing A1, A3 - all models
Operation with an ECG Simulator A1 - M3923A, M3924A
A3 - all models
Operation with a Respiration Simulator A3 - all models
Verification of Pneumatic System A1, A3 - all models
Operation with a Temperature Simulator A1 - M3922A, M3924A
A3 - all models
General Operation A1, A3 - all models
SWITCH POSITION
RATE 38
LIGHT LOW
MODULATION OFF
RCAL/MODE RCAL 63/LOCAL