Power-up Defaults Menu And Diagnostic Mode 37
4
Power-up Defaults Menu And Diagnostic
Mode
Introduction
This section discusses use of the Power-up Defaults Menu to configure
power-on default settings, and the Diagnostic Mode to obtain service-related
information about the monitor.
Power-up Defaults Menu
The purpose of the Power-up Defaults Menu is to allow the authorized user to
create a power-up default for each setting in the monitor. Power-up defaults
are the settings in effect each time the monitor is powered on. Once the
Power-up Defaults Menu is entered, physiological monitoring is terminated.
The screen layouts do not display any information associated with normal
monitoring operation.
Use the following procedure to configure the power-up default settings for the
monitor:
1. Set the monitor to normal monitoring mode.
2. Adjust each accessible setting on the monitor as desired.
Note
Use the techniques described in the User Guide. Such settings
include alarm limits, choice of display type for the graphic
frames, and ECG lead select.
3. Use the wheel to invoke the Set-up Menu (choose the screwdriver icon
found along the bottom of the display).
4. Select menu item
Enter Power-Up Defaults Menu. Once
selected, a pop-up box will appear with the text
Enter 3-Digit
Passcode.
5. Use the wheel to enter the passcode,
2 1 5.
Note
This passcode is set at the factory and cannot be changed.
6. The Power-up Defaults Menu is now present. The available menu items are
explained below. Make changes to these menu items as desired.

Table of Contents