Diagnostic Mode
Power-up Defaults Menu And Diagnostic Mode 41
error code. Error codes cannot be changed or reset in this screen. When in the Error
Code screen, the
Return item is always highlighted; a press of the wheel returns the
user to the Diagnostic Menu. Rotating the wheel while in the Error Code screen has no
effect.
Refer to “Error Codes”, Chapter 5, for more detail on error codes.
System Information
This screen displays several system-related items:
Monitor On-time: Displays the number of hours, rounded to the
nearest hour, that the Main PCB has been
operational. This value cannot be reset.
Note The Monitor On-time, Backlight On-time, and Battery Deep Discharge
values are stored in nonvolatile memory. When a new Main PCB is
installed, this value is reset to zero.
Backlight On-time: Displays the number of hours, rounded to the
nearest hour, that the LCD Backlight has been
operational. This value can be reset to zero, for
instance at the time when a technician changes the
backlight or installs a new LCD.
Recorder On-time: Displays the number of hours, rounded to the
nearest hour, that the A3 Recorder has been
SYSTEM INFORMATION
Monitor On-Time 1563
Backlight On-Time 871
Recorder On-Time
1
1. A3 only
37
Battery Deep Discharges 152
System Software Version V1.02
SpO
2
Software Version V1.2.0
Return