42
SimplyGo User Manual
*Based on an atmospheric pressure of 14�7 psia (101 kPa) at 70° F (21° C)
** Per ISO 80601-2-69, the maximum Sound Pressure Level is 58 ± 3 dBA and the maximum
Sound Power Level is 66 ± 3 dBA at a continuous setting of 2 LPM� The maximum Sound Pressure
Level is 51 ± 2 dBA and the maximum Sound Power Level is 59 ± 2 at Nominal Pulse Mode Setting
of 6 and 20 BPM�
SimplyGo Flow Settings and Pulse Volumes
Pulse Mode
Settings
1�0 1�5 2�0 2�5 3�0 3�5 4�0 4�5 5�0 5�5 6�0
Breath
Rate
Pulse Volumes (mL)
15 11�5 17�2 24�0 30�0 36�0 42�0 48�0 54�0 60�0 66�0 72�0
20 11�5 17�2 24�0 30�0 36�0 42�0 48�0 53�8 59�5 65�5 71�5
25 11�4 17�1 23�9 29�8 35�6 39�0 42�3 44�0 45�7 51�0 56�2
30 11�4 17�0 22�6 25�8 28�9 30�4 31�8 33�3 34�7 38�8 42�9
35 11�3 14�2 17�1 19�6 22�0 23�3 24�6 25�9 27�1 30�3 33�4
40 10�9 11�9 12�9 14�9 16�8 17�9 19�0 20�0 20�9 23�3 25�7
+/-15% at STPD*
+/-25% over the rated environmental range
Average of 30 consecutive pulses with 1�5 m length tubing
Audio Alarm 50 decibels (nominal) at 39 in� (1 m)
Outlet Pressure 6�4 psig maximum (44 kPa)
Expected Service
Life of Device and
Accessories
5 years