43
SimplyGo User Manual
English
Sleep Mode
Settings
1�0 1�5 2�0 2�5 3�0 3�5 4�0 4�5 5�0 5�5 6�0
Breath
Rate
Pulse Volumes (mL)
15 16�7 25�0 33�3 41�7 50�0 58�3 66�7 75�0 83�3 91�7 100�0
20 12�5 18�2 23�9 29�8 35�7 41�6 47�4 53�2 59�0 64�7 70�4
25 8�7 13�2 17�6 22�0 26�4 30�8 35�2 39�5 43�7 47�9 52�1
30 6�4 9�9 13�3 16�7 20�0 23�5 26�9 30�2 33�5 36�7 39�8
35 4�7 7�5 10�2 12�9 15�5 18�2 20�9 23�6 26�2 28�6 31�0
40 3�3 5�6 7�8 9�8 11�8 14�0 16�1 18�2 20�3 22�2 24�0
+/-15% at STPD*
+/-25% over the rated environmental range
Average of 30 consecutive pulses with 1�5 m length tubing
Continuous Mode
Settings
0�5 1�0 1�5 2�0
Minute Volumes (mL/min)
500�0 1000�0 1500�0 2000�0
+/-15% at STPD*
+/-25% over the rated environmental range
Three minute running average
*STPD is 101�3 kPa at an operating temperature of 20 °C, dry�
Note: Max recommended ow is 2 LPM (at nominal outlet pressures of 0 and 7 kPa)�