EasyManua.ls Logo

Philips SimplyGo - Technical Specifications; Operating and Storage Conditions; Oxygen Delivery Specifications

Philips SimplyGo
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
SimplyGo User Manual
Oxygen
Concentration*
87%-96% at all settings
FlowSettingsand
Pulse Volumes
Pulse Mode
1 = 12 ml; 1½ = 18 ml; 2 = 24 ml; 2½ = 30 ml;
3 = 36 ml; 3½ = 42 ml; 4 = 48 ml; 4½ = 54 ml;
5 = 60 ml; 5½ = 66 ml; and 6 = 72 ml;
+/- 15% or 4 ml, whichever is greater (Average of 20
consecutive breaths)
up to a max of 2000 ml/min +/- 200 ml
Sleep Mode
Variable pulse volumes, based on breath rate, to
maintain a constant minute volume per setting�
1 = 250 ml; 1½ = 375 ml; 2 = 500 ml; 2½ = 625 ml;
3 = 750 ml; 3½ = 875 ml; 4 = 1000 ml; 4½ = 1125 ml;
5 = 1250 ml; 5½ = 1375 ml; and 6 = 1500 ml
+/- 15% or 60 ml, whichever is greater (Average of
20 consecutive breaths)
Continuous Mode
½ = 500 ml/min; 1 = 1000 ml/min;
1½ = 1500 ml/min; 2 = 2000 ml/min;
+/- 10% or 100 ml/min, whichever is greater
(3 minute running average)
Note: Max recommended ow is 2 LPM (at nominal
outlet pressures of 0 and 7 kPa)�
Specications
Operating
Conditions
Operational temperature:
41°Fto104°F(5°Cto40°C)
Relative humidity: 15% to 95%
Altitude: up to 10,000 ft� (3048 m�)
Storage and Transport
Conditions
-4°Fto140°F(-20°Cto60°C)-unitonly
Relative humidity: up to 95%, non condensing
SimplyGo Oxygen Concentrator