EMC Information
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
SimplyGo is intended for use in the electromagnetic environment specied below� The user of SimplyGo should assure that it is
used in such an environment�
Emissions Test Compliance Electromagnetic Environment - Guidance
RFEmissions
CISPR 11
Group 1 TheDeviceusesRFenergyonlyforitsinternal
function.Therefore,itsRFemissionsarevery
low and are not likely to cause any interference
in nearby electronic equipment�
RFEmissions
CISPR 11
Class B The Device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes�
Harmonic Emissions
IEC 61000-3-2
Class A
VoltageFluctuations/FlickerEmissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
SimplyGo is intended for use in the electromagnetic environment specied below� The user of SimplyGo should assure that it is
used in such an environment�
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment - Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV Contact
±8 kV Air
±6 kV Contact
±8 kV Air
Floorsshouldbewood,concreteorceramictile.
If oors are covered with synthetic material, the
relative humidity should be at least 30%�
ElectricalFast
Transient/Burst
IEC 61000-4-4
±2 kV for Power Supply
Lines
±1 kV for Input/Output
Lines
±2 kV for Power Supply
Lines
±1 kV for Input/Output
Lines
Mains power quality should be that of a typical
home or hospital environment�
Surge
IEC 61000-4-5
±1 kV Line to Line
±2 kV Line to Ground
±1 kV Line to Line
±2 kV Line to Ground
Mains power quality should be that of a typical
home or hospital environment�
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Input
Lines
IEC 61000-4-11
<5% U
T
(>95% Dip in U
T
)
for 0�5 Cycle
40% U
T
(60% Dip in U
T
)
for 5 Cycles
70% U
T
(30% Dip in U
T
)
for 25 Cycles
<5% U
T
(>95% Dip in U
T
)
for 5 Seconds
<5 % U
T
(>95% Dip in U
T
)
for 0�5 Cycle
40 % U
T
(60% Dip in U
T
) for
5 Cycles
70 % U
T
(30% Dip in U
T
) for
25 Cycles
<5 % U
T
(>95% Dip in U
T
)
for 5 Seconds
Mains power quality should be that of a typical
home or hospital environment� If the user of the
Device required continued operation during
power mains interruptions, it is recommended
that the Device be powered from an
uninterruptible power supply or battery�
PowerFrequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds should be at
levels characteristic of a typical location in a
typical home or hospital environment�
Note: U
T
is the A�C� mains voltage prior to application of the test level�