14
SimplyGo User Manual
Device Warm-Up Breathing
When you turn on your SimplyGo device, it will sense if you are breathing
from it� If you are not yet breathing through the cannula, SimplyGo will begin
to pulse automatically about once every ve secondsThe only purpose of
this device warm-up breathing is to help the device warm up faster� As soon
as you begin breathing through the cannula, the device will begin delivering
pulses based on your breathing� Device warm-up breathing occurs only when
you rst turn on the deviceYou may begin breathing from the SimplyGo at
any time� If the device does not sense a breath after two minutes, a No Breath
alarm will sound�
Operating Modes
This section describes additional operating screens on the SimplyGo device
1. Pulse Mode Screen
When the device is in the Pulse operating mode, it detects when the user
begins to take a breath and then delivers a pulsed volume of oxygen
determined by the setting number, during inhalation� If no breath is de-
tected for a period of time, the system automatically delivers the pulsed
volume of oxygen determined by the setting number at a fixed rate of
12 breaths per minute� If no breath is detected for 2 minutes, the device
signals an alarm and shuts down after 30 minutes
2. Continuous Flow Mode Screen
InContinuousFlowoperatingmode,aconstantflowofoxygenissup-
plied based on the setting number (in liters per minute):
• .5LPM
• 1LPM
• 1.5LPM
• 2.0LPM