19
SimplyGo User Manual
Alarm Indicators and Screen Symbols
Visual, Audio
Indicators
Description What To Do
3 short
beeps
+
1 long beep
No Breath Alarm
This alarm occurs
when a breath is not
detected for a period
of 2 minutes or more
It can be silenced by
pressing the Alarm
symbol� This alarm
becomes silent as soon
as a breath is detected�
If no breath is detected
after approximately
30 minutes, the device
shuts down to conserve
power� (This alarm is
inactive for ten minutes
after startup�)
Check the connection
from the cannula to
the device� Make sure
the nasal cannula is
properly positioned
on your face and that
you are breathing
through your nose�
Make sure the
cannula tubing is not
kinked or obstructed�
3 long beeps
+
1 short beep
Low Oxygen
Concentration Alarm
This alarm occurs when
the device is delivering
a lower concentration of
oxygen than specied�
This alarm can be
silenced by pressing the
Alarm symbol�
Change to another
source of oxygen
and contact your
equipment provider

Other manuals for Philips SimplyGo