20
SimplyGo User Manual
Visual, Audio
Indicators
Description What To Do
1 long beep
after 2 minutes
High Breath Rate Alarm
This alarm indicates that
the users breath rate is
exceeding the capacity
of the device� The device
is still working properly
and is still providing
oxygen at the rate
specied for the setting
The beeps can be
silenced by pressing the
Alarm symbol�
The indicator resets
itself when the breath
rate is reduced� If this
indicator regularly
occurs, contact your
home care provider
3 short beeps
+
3 long beeps
Technical Fault Alarm
This alarm occurs when:
a) the battery runs out
or
b) the device has a
general malfunction
and the device is no
longer operating
properly
The device may or
may not shut down
depending on the
severity of the alarm�
This alarm can be
silenced if the
symbol appears
on-screen�
Recharge the battery
and try to restart the
device�
If this alarm occurs
again, change to
another source of
oxygen and contact
your home care
provider