21
SimplyGo User Manual
Visual, Audio
Indicators
Description What To Do
1 short beep
Low Battery Alarm
This alarm occurs
when approximately
17 minutes of
battery life remains�
Remaining battery life
is dependent on your
device settings and your
activity levelThe empty
battery symbol ashes
on-screen�
Replace the battery
or connect to a power
source
(Press the Alarm
symbol to silence the
alarm�)
3 long beeps
+
1 short beep
No Flow Alarm
This alarm occurs when
the device detects there
is no oxygen owing in
the patient cannula�
Check the cannula
for kinks or other
obstructions that are
stopping oxygen ow
through it�
Alternating short
and long beeps
Depleted Battery
Alarm
This alarm occurs when
approximately two
minutes of battery
life remains� The
battery symbol ashes
on-screen�
Replace or recharge
battery
(Press Alarm symbol
to silence the alarm�)
Alarm Silence Symbol
This symbol appears
when you press the
Alarm Silence button
to silence the audible
alarm (beeps)�
No further action is
needed�