AUTOMATIC EXPOSURE CONTROL (AEC) (OPTIONAL)
38 Planmeca Proline EC Panoramic X-ray
User's manual
Adjustment range of AEC exposure parameter
Child:
the range varies from -4 to +3 units compared
to the original values.
Small or average patient:
the range varies from -5 to
+5 units compared to the original values.
Large patient:
the range varies from -5 to +6 units
compared to the original values.
One unit corresponds to 1 mA or 2 kV, i.e. the maxi-
mum increase of the exposure parameter for example
for child exposure can be 3 mA and 6 kV or
2 mA and 2 kV.
Exiting the AEC mode
Exit the AEC mode by pressing and holding down the
auto key until the indicator light goes off.