Planmeca ProSensor® HD intraoral sensor
Remember to wear protective gloves
Switch the device off before cleaning and disinfection
Approved disinfectants for wiping surfaces
• Dürr: FD 322, FD 333
• Metrex Research: CaviWipes
• Johnson & Johnson: Cidex Opa
Approved immersion disinfectant
• Johnson & Johnson: Cidex Opa
Sensor and cable
Use a new disposable hygienic sheath every time you use a sensor
To disinfect the sensor, wipe its surface with a soft cloth dampened with a disinfectant solution
The sensor casing can also be immersed in disinfectant solution
The sensor connector can be cleaned using a soft cloth
Do not leave the sensor in disinfectant overnight
The magnetic connector of the sensor cable should not be soaked
Do not sterilise the sensor or cable in an autoclave or UV oven
Sensor control box
Remember to switch the device off before cleaning
Clean the control box with a soft cloth dampened with a mild cleaning solution
Do not disinfect the control box