Planmeca ProScanner® 2.0 imaging plate scanner
Remember to wear protective gloves
Switch the scanner off before cleaning and disinfection
Imaging plate scanner
Clean and remove any soiling with a soft, lint-free cloth dampened with cold tap water
Disinfect the surfaces with a disinfection wipe, or use disinfectant on a soft, lint-free cloth
Approved disinfectants
• Dürr: FD 322, FD 333, FD 350, FD 366 sensitive
Make sure no liquids get inside the scanner, as they could damage it
Do not spray the scanner
Imaging plates
Only use wipes intended for cleaning imaging plates
• e.g. Dürr Dental imaging plate cleaning wipes
Remove resistant or dried on impurities and disinfect with an imaging plate cleaning wipe
Use a new disposable light protection cover every time you use an imaging plate
Allow the plates to dry completely before using them
Do not steam sterilise the plates – heat and humidity will damage them
Do not immerse the plates in disinfectant
Cleaning and disinfectant wipes are unsuitable for cleaning imaging plates
Light protection covers
Disinfect the covers using disinfectant before and after placement
Approved disinfectant wipes:
• Dürr: FD 333 forte (virucidal), FD 350 (limited virucidal activity), FD 322 premium (limited
virucidal activity)
Allow the covers to dry completely before using them