04/2020
www.planmeca.com
Cartridges
󰣑󰣒 Remove any impurities with a soft, lint-free cloth that has been dampened with cold tap water
󰣑󰣒 Clean and disinfect with a disinfectant wipe, or with a soft, lint-free cloth dampened with disinfectant
󰣑󰣒 Alternatively, the cartridges can be immersed in disinfectant
󰣑󰣒 Approved disinfectants
Dürr: ID 213, ID 212, ID 212 forte
󰌦 Allow the cartridges to dry completely before using them
 Do not thermally disinfect or steam sterilise the cartridges, as heat will damage them
Imaging plate storage box
󰣑󰣒 Use a soft, lint-free cloth dampened with cold tap water to clean the surfaces of the box
󰣑󰣒 Use a disinfectant wipe to disinfect the surfaces,
󰣑󰣒 Alternatively, the image plate storage tray can be treated in a thermal disinfector or steam steriliser
󰣑󰣒 Approved disinfectants
For the storage box
o Dürr: FD 366 sensitive
For the image plate storage tray
o Dürr: FD 350, FD 366 sensitive
 If a thermal disinfector or steam steriliser is used, do not exceed the temperature of 134 °C (273 °F)

Other manuals for Planmeca ProX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProX and is the answer not in the manual?

Planmeca ProX Specifications

General IconGeneral
TypeIntraoral X-ray
Tube current7 mA
Imaging TechnologyDigital
Focal spot0.4 mm
X-ray generatorHigh-frequency
Power Requirements100–240 V AC, 50/60 Hz
Tube voltage60-70 kV

Related product manuals