Planmeca ProMax® and Planmeca Viso™ extraoral
X-ray units
Remember to wear protective gloves
Switch the unit off before cleaning and disinfection
Approved cleaning agents
• Alpro Medical: CleanWipes, IC-100, MinutenSpray-classic
• Clinell: Clinell Universal Wipes
• Clinitex: R515 Detergent Multi-surface wipes
• EcoLab: Actichlor Plus
• SciCan: Optim Blue Wipes
Approved surface disinfectants
• Alpro Medical: MinutenSpray-classic
• Antibac: Antibac 75%
• CCS Healtcare: Dax Extra, Dax 70+
• Chemi-Pharm AS / Plandent: Orbis Surface Disinfectant
• Clinell: Clinell Universal Wipes
• Dürr: FD 312(*)
• Ecolab: Actochlor Plus
• SciCan: Optim Blue Wipes
(*) Note: not suitable for Planmeca Viso’s head support.
Do not apply sprays, liquids or foams directly onto any surface – use a clean, damp cloth instead
X-ray unit patient supports, handles and touch screens
Change the bite stick to a clean one for each patient or use a disposable protective shield
Wipe all the parts after each patient using a surface disinfectant approved by Planmeca
For stains and other impurities, use a Planmeca approved cleaning agent when needed
Some support parts can be autoclaved at 134 °C (273 °F): a 3–4 minute cycle for porous loads and
instruments – see further details in the user’s manual
Other surfaces of X-ray units
Wipe parts regularly using a surface disinfectant approved by Planmeca
For stains and other impurities, use a Planmeca approved cleaning agent when needed