17
Possible values
Initial value Increment Lower value Upper value Remarks
Measuring range 0…2 ppm 0…0.5 ppm
0…2 ppm
0…5 ppm
0…10 ppm
0…20 ppm
0…50 ppm
0…100 ppm
D1C2-Cl-030-GB
measuring range
setting ?
measuring range
0…2.00 ppm
Permanent display 1
Complete Operating Menu / Description
Measuring range
Error message Condition Comment
Buffer distance too small Buffer < pH 2 During calibration procedure:
Recalibrate buffer 2!
Return to permanent display:
Zero point low < -60 mV Basic metering load Warning, old zero point and
slope retained
Zero point high > +60 mV " "
Slope low < 45 mV/pH " "
Slop high > 65 mV/pH " "
Measured value pH unstable "
Measurde value °C unstable "
IMPORTANT
The chlorine sensor must be recalibrated and the settings checked in all menus after
changing the range allocation!
Measured value
IMPORTANT
The chlorine sensor must be recalibrated and the settings checked in all menus after
changing the range allocation!
D1C2-Cl-020-GB
Access to all setting menus can be blocked with an access code !
Standard signal input mA
measured value
range adjustment
4 mA
= 0.00 ppm
20 mA =
2.00 ppm
measured value
range adjustment
measured value
checkout time
measured value
off
meas. value
setting ?
BA_DM_173_03_09_GB.p65 30.03.2009, 12:10 Uhr17